Code 1001-112

Troubleshooting

Code 1001-112

Category: Troubleshooting

Symptom

The following code was reported in Pulsewave® Software: 1001-112: Systolic Time (STI) and Diastolic Time (DTI) could not be calculated.

Applies To

Pulsewave® BP+

Explanation

The Target pressure high was not high enough. If the Target pressure high is not high enough, data required to measure STI and DTI may not be captured. These results are not given with this code. However, other results may be provided.

Troubleshooting

If you are the only person using your account to submit readings:

  1. Confirm that Target pressure preference is set to Auto.*
  2. Take another reading.

If more than one person submits readings using your account:

  1. Confirm that Target pressure preference is set to Fixed.*
  2. Take another reading.
  3. If the code repeats, then set the Target pressure preference to Manual.*
  4. Set the Target pressure high to “280,” and the Target pressure low to “45.”
  5. Take another reading.

* To learn more about Target pressure preference, please read the instructional article “Using Reading Mode.”