Code 1001-20
Category: Troubleshooting
Symptom
The following code was reported in Pulsewave® Software: 1001-20: The target pressure range was too small.
Applies To
Pulsewave® Software
Explanation
An insufficient number of pulses were captured during the reading because the target pressure range set for the cuff was too small. If the range between Target pressure high and Target pressure low is too small, the amount of data captured is too limited for analysis to provide results. Results are not provided with this code.
Troubleshooting
If you are the only person using your account to submit readings:
- Confirm that Target pressure preference is set to Auto.*
- Take another reading.
If more than one person submits readings using your account:
- Confirm that Target pressure preference is set to Fixed.*
- Take another reading.
* To learn more about Target pressure preference, please read the instructional article “Using Reading Mode.”