Code 2000-0

Troubleshooting

Code 2000-0

Category: Troubleshooting

Symptom

The following code was reported in Pulsewave® Software: 2000-0: The cuff was too loose or not attached correctly.

Applies To

Pulsewave® Device

Explanation

This code is reported when the cuff is loose enough or incorrectly placed that it inflates and reaches the maximum time to inflation before the Target pressure high was reached.

Troubleshooting

  1. Confirm cuff placement on the wrist (preferably the left wrist)
  2. Tighten the cuff and take another reading

Still need help?

Customer Support